Arbitrary Waveform Generator (AWG): Difference between revisions
Line 51: | Line 51: | ||
<pre> | <pre> | ||
Value;Field;Description | Value;Field;Description | ||
8;Samples;The number of samples to use (later samples are ignored) | |||
244;SampleStretch;The number of times to repeat a sample | 244;SampleStretch;The number of times to repeat a sample | ||
0,1;BeginData;Data begins here | 0,1;BeginData;Data begins here | ||
0,2;; | 0,2;; | ||
0,3;; | 0,3;; |
Revision as of 09:25, 2 March 2015
The SmartScope has an Arbitrary waveform generator, capable of generating signals between the [0V, 3.3V] voltage range at a sample rate of 100MS/s.
AWG pin location
The signal generated by the AWG is presented on the 3rd-left pin on the bottom row of the AUX connector, as shown in the following image:
Please keep in mind that you always should bridge 2 wires between 2 separate devices. In this case:
- The AWG output signal
- The ground, so both devices have the same reference voltage. (0V means the same on both devices)
In the image above, you can see the AWG pin is surrounded by 2 ground pins, either of which you can use to connect to the other device.
How the AWG works
The AWG is driven by a 100MHz clock, reading out a memory containing up to 2048 samples.
Parameter | Range | Description |
---|---|---|
Samples | 1-2048 | The number of samples used. The AWG loops on these samples. (< 0.0.8.4 had a minimum of 128 samples) |
SampleStretch | 0-255 | The number of cycles to repeat each sample |
BeginData | Field to indicate that from here on the sample data follows |
Configuring the AWG using pre-defined waves
Configuring the AWG using csv files
IMPORTANT: The app expects the CSV file to use a semicolon (;) as a field separator and a comma (,) as decimal symbol. Download one of the samples below to make sure your CSV works. NOTE: The CSV file may contain more than the number of samples specified in the parameters. Further samples will simply be ignored
NOTE: This only works with a SmartScope connected. Otherwise, the side menu won't contain an AWG item If you haven't used dropbox with the SmartScope
- Tap sidemenu > AWG > Upload from dropbox
- The app will tell you it doesn't have permission to dropbox and ask for it by sending you off to the dropbox website
- Grant access and return to the app
- The app now creates the AWG folder and will inform you that this new folder is empty.
- Generate a CSV file using the AWG excel worksheet. A sample CSV can be found for a sine and block wave.
- Drop your CSV file in the AWG folder (<dropbox>/Apps/LabNation SmartScope/AWG) using a file manager
- In the app, tap sidemenu > AWG > Upload from dropbox
- You should now be able to choose the CSV file
Sample CSV
Value;Field;Description 8;Samples;The number of samples to use (later samples are ignored) 244;SampleStretch;The number of times to repeat a sample 0,1;BeginData;Data begins here 0,2;; 0,3;; 0,4;; 0,3;; 0,2;; 0,1;; 0;;