Digital Waveform Generator

From LabNation Wiki
Revision as of 20:17, 7 August 2016 by Riemerg (talk | contribs)
Jump to navigation Jump to search

NOTE: the current version of software (v0.0.8) does not yet support configuring the digital output channels automatically.

The SmartScope comes with 4 dedicated digital output channels, each capable of generating up to 100MS/s.
This allows square waves of up to 50MHz to be generated.

Output pin locations

The 4 output channels are located on the AUX port, at the back of the SmartScope.
DWGoutput.png

Output voltage levels

As output voltage levels, you can select between 2 standards:

  • 0V - 3.3V
  • 0V - 5.0V

These voltages are selectable for an entire waveform and for all 4 channels together, not per sample nor per channel. In order to select the voltage, go to Menu -> Generator -> Digital -> Voltage level

Define arbitrary waveforms in Excel (.csv file upload)

Parameter Range Description
DataIsBytes 0 or 1 Indicates if the data should be interpreted as bytes (for the 4 digital outputs). If '0', it will be interpreted as voltages (for the analog output).
Samples 1-2048 The number of samples used. The AWG loops on these samples. (< 0.0.8.4 had a minimum of 128 samples)
SampleStretch 0-255 The number of cycles to repeat each sample
BeginData Field to indicate that from here on the sample voltage levels follow

Each row defines a sample as byte value, where the 4 MSB define the values of the 4 output pins.

CsvValue Output3 Output2 Output1 Output0
0 0 0 0 0
1 0 0 0 1
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
10 1 0 1 0
etc etc etc etc etc


Sample digital output CSV

Value;Field;Description
1;DataIsBytes;Set to 1 to interpret the data as bytes, otherwise interpreted as voltages
8;Samples;The number of samples to use (later samples are ignored)
244;SampleStretch;The number of times to repeat a sample
0;BeginData;Data begins here
1;;
2;;
4;;
8;;
4;;
2;;
1;;